Modification of electrodes with N-and S-doped Carbon dots.
Evaluation of the electrochemical response
Marta Bonet-San-Emeterioa, Manuel Algarrab,*, Marijana Petković,b Manel del Vallea,*
a
Sensors and Biosensors group, Department of Chemistry, Universitat Autònoma de
Barcelona, 08193 Bellaterra, Barcelona, Spain.
b
CQM-Centro de Química da Madeira, Universidade da Madeira, Campus da Penteada,
9020-105 Funchal, Portugal.
Abstract
Nitrogen and sulphur-doped Carbons Dots (N-CDs and S-CDs) were synthesized by a
hydrothermal method and incorporated as surface electrode modifiers to evaluate their
properties for electrochemical sensing. The first task was to characterize the synthesized
materials, for which different spectroscopies, scanning microscopes, mass spectrometry
and elementary analysis were performed. Next, a glassy carbon electrode (GCE) was
surface-modified with the doped CDs and applied to check the electrochemical signal of
different organic compounds corresponding to different families. Water solubility of the
doped carbon dots forced us to incorporate them in a graphite-polystyrene ink to
accomplish the modification of electrodes. This modification needed a first activation to
obtain a properly conductive surface. The organic compounds examined were salicylic
acid, cysteine and ascorbic acid. The modified GCEs exhibited an enhanced sensitivity,
probably caused by the increase of active surface, but in addition, signals of salicylic acid
were shifted to lower potentials, what is a proof of the increase of the heterogeneous
electron transfer rate, and a demonstration of an enhanced catalytic response.
Keywords
Modified electrode; Voltammetry, N-doped carbon dots; S-doped carbon dots,
Nanomaterial; Electrochemical sensor.
1
1 Introduction
Carbon quantum dots, or simply, carbon dots (CDs) are an emerging new nanomaterial
made essentially of carbon that can be considered small graphene nanoplatelets, or
other small carbon aggregates [1]. Typical sizes involved are in the range of 10 nm, and
in some aspects, may overcome problems assigned to conventional metal-
semiconductor quantum dots (QDs) [2]. Among others, some advantages of this new
carbon nanomaterial are their stability, water solubility, low toxicity, high fluorescence,
biocompatibility and easy chemistry for conjugation or functionalization [3].
Their most interesting analytical property is their photoluminescence in the VIS range,
which has made this nanomaterial to be exhaustively studied in the development of new
fluorometric or fluorescence quenching methods of analysis. This is so, because
photoluminescence and reactivity of CDs may be tuneable by changing their size,
depending on the synthesis method, and by introducing various functional groups at the
graphitic sheet edges [4]. For all these, CDs are to be considered as promising
applications in bio-sensing, cellular imaging, drug delivery and catalysis [5,6]. Inspired
by the variety of their usage, there is a growing interest in the synthesis of new carbon
dots, where they are in the forefront of new environmentally compatible, or green
analytical methodologies [7].
Most of the articles present in the literature related to CDs exploit their potential
alternative to conventional semiconductor QDs as fluorescent material for imaging [8-
10]. In this area, CDs are playing up their importance in a number of applications, thanks
to their inherent properties of hydrophilicity, stable luminescence, resistance to
photobleaching, chemical stability, low cost and simple synthesis and functionalization
2
[11-15]; all these properties make them appropriate and/or ideal nanomaterials for
selective imaging, catalysis or bactericide effects, among others existing research fields
[3,4, 16-22]. Nonetheless, CDs are gaining importance in new fields as in sensing and
biosensing. For example, Praneerad et al. describe a CD fluorescent paper-based sensor
for acetic acid vapour and Cu2+ detection [23]. Thanks to their semiconducting
properties, CDs have also been successfully applied as a part of binary matrices for
matrix-assisted laser desorption/ionization (MALDI) mass spectrometry of low mass
biologically relevant molecules [24, 25], enhancing the signal intensity.
Almost practically all the initial research was related with purely carbon-containing CDs,
soon it was realized that their doping with some elements such as nitrogen, sulphur,
phosphorous or boron, or their combinations, could be used to tune their properties.
Initially, the focus was put on the fluorescence emission, which could be altered from
blue to green [26-28]. The nitrogen or sulphur atoms, which especially populate the
surface of the CDs, may act as electron-accepting donors increasing their power, in
terms of sensitivity, to interact with different analytical species. In a different
application, nitrogen- and sulphur- doped carbon dots, sprayed over the tissue sections,
lowered down the sensitivity of small molecules’ detection by MALDI mass spectrometry
imaging to the picomolar range, as demonstrated in recent work of Lin et al., thus
becoming an important auxiliary agent in (bio)analytics [29].
Doped CDs may be prepared by different approaches, from small organic precursors,
being microwave [30], pyrolysis [31] and hydrothermal [32-36] the preferred methods.
Three main outcomes can be obtained: the surface modification/ functionalization,
doping with heteroatoms and therefore the control of crystallinity [37].
3
Concerning the use of CDs in electroanalysis, much less research has been done in this
area of sensing [38, 39]. This alternate use of CDs can be directly aimed to modify
electrodes, alone, or in combination with other compounds; this strategy is inspired in
the different uses reported for graphene, a greatly used precedent material for
electrochemical sensors. One of the first works reported was a communication from
Nguyen et al. in which they used a CDs-modified glassy carbon electrode for the
determination of etoposide [40]. From 2018, there is also a report from our group, in
which it was developed a totally functional sensor modified with CDs for the
determination of dopamine and uric acid [41]. Other species determined with similar
devices have been hydroquinone and catechol [42], adrenaline [43] and catecholamine
type neurotransmitters [44]. As it is clear, CDs have been more widely applied in the
electrochemical analysis of organic molecules than in other instances. In general, CDs
may act in these oxidative determinations as multivalent redox species, able to facilitate
electron transfer and to bring catalytic advantages, as it is the case with noble metal
nanoparticles [39].
With respect to the use of doped CDs, even less articles can be found in the literature.
Among the few, there is the electrocatalytic sensor based on boron-doped CDs for the
detection of haemoglobin [45] or one sensor modified by electrodeposition with N-
carbon quantum dots and copper oxide for the selective detection of aspirin [46].
Another recent work describes the use of N-doped CDs for detection of hydrogen
peroxide and paracetamol [47]. In the present work we first introduce CDs doped with
sulphur and nitrogen and their use as modifiers on glassy carbon electrode for the
detection of salicylic acid, cysteine and ascorbic acid, where we have compared their
electroanalytical behaviour, also in front of pristine CDs.
4
2 Experimental
2.1. Chemicals and reagents
All analytes used herein were of analytical reagent grade. The solutions were prepared
using deionized water from a Milli-Q system (Millipore, Billerica, Ma, USA).
Polyvinylpyrrolidone (PVP, average mol. 40000) and Poly(4-styrenesulphonic acid)
sodium salt (PSS, average mol. 1000000) used for N-doped and S-doped CDs synthesis,
respectively, were from Merck (Darmstadt, Germany). The analytes for electrochemical
measurements, ascorbic acid, salicylic acid and cysteine were from Panreac Química
(Barcelona, Spain), Merck (Darmstadt, Germany) and Acros Organics (Geel, Belgium),
respectively.
2.2 Carbon dots synthesis and modification
Methods described in the literature, with small modifications to adapt them to our
laboratory, were used to obtain N-CDs and S-CDs [22]. Briefly, PVP and PSS (2 g), as start
reagents for N-CDs and S-CDs respectively, were dissolved in H2O (50 mL), the obtained
solutions were transferred into a 50 mL teflon lined steel reactor and made react in an
oven at 180°C for 3 h. The resulting brown aqueous solution showed a green
fluorescence under UV light. For the purification of both N- and S-CDs, the obtained
mixtures were centrifuged and dialyzed overnight (0.1-0.5 kDa, Biotech). Finally, both
solutions were lyophilized yielding the solid N-CDs and S-CDs nanoparticles,
respectively.
2.3 Characterization of doped CDs
5
CHNS determinations of N-CDs and S-CDs were performed by using a Thermo Scientific
FLASH 2000 elemental analyser. Samples were weighed in a tin capsule and introduced
into the combustion reactor via the Thermo Scientific MAS 200R autosampler, in all
cases ensuring the proper amount of oxygen was present. After combustion at 1200°C,
the resultant gases were carried by a helium flow to a catalytic layer filled with copper,
then swept through a GC column that separated the combustion gases, and finally
detected by a thermal conductivity detector (TCD). XPS studies were carried out on a
Physical Electronics PHI Versa Probe II spectrometer using monochromatic
Al-Kradiation (49.1 W, 15 kV and 1486.6 eV) for analysing the core-level signals of the
elements of interest with a hemispherical multichannel analyser. The sample spectra
were recorded with a constant pass energy value at 29.35 eV, using a 200 μm diameter
circular analysis area. The X-ray photoelectron spectra obtained were analysed using PHI
Smart software and processed using the Multi Pak 9.3 package. The binding energy
values were referenced to adventitious carbon C 1s signal (284.8 eV). Shirley-type
background and Gauss-Lorentz peak curves were used to fit the data and determine the
binding energies. Scanning electron microscopy (SEM) images were obtained using a
scanning electron microscope EVO®MA10 (Zeiss, Germany) operated at 20 kV.
CDs were also characterized by matrix-assisted laser desorption/ionization time of flight
mass spectrometry (MALDI TOF MS) on the Autoflex Max Device (Bruker, Bremen,
Germany), equipped with the SmartBeam® laser emitting at 355 nm. Maximum laser
frequency is 2 kHz, and all spectra were acquired using reflector detector and delayed
extraction time 120 ns. CDs were dissolved in water at concentration of 1 mg·mL-1, and
small volume (0.5 µL) was applied on the stainless steel MALDI target and left at the
room temperature to crystallize.
6
2.4 Electrochemical characterization
The synthesized carbon dots were immobilized on glassy carbon electrode (GCE) using
graphite-based inks. These inks were prepared using polystyrene as polymeric part and
mesitylene as solvent. Concretely the mixture contained the following mass fractions:
58% of graphite, 32% of powdered polystyrene and a 10% of modifier, in this case the
CDs. Once the mesitylene was added to the mixture it was homogenized by stirring it for
2 hours and placing it in an ultrasonicating bath for 2 minutes. A GCE, previously polished
using wet alumina powder from 1.0 to 0.05 µm, was modified with 5 µL of ink and dried
in a 40oC heater. Then the surface was electrochemically activated using a 50 mM H2O2
solution (Merck, Darmstadt, Germany) in a phosphate buffer solution. The electrode
was activated in the H2O2 solution by applying 10 cyclic voltammetry (CV) cycles in a
wide potential range (from +1.50 V to -1.50 V) at 0.5 V·s-1 [48,49]. Voltammetric
measurements were carried out with a MultiEmStat multichannel potentiostat
controlled by MultiTrace software (both from PalmSens BV, Randhoeve, The
Netherlands); a typical three electrode cell was used, formed by a combination
reference electrode of Ag/AgCl (3 M KCl) and Pt counter electrode, respectively, and the
working electrode designed herein. The data was recorded using the current between
-1.0 and +1.5 V with a step potential of 0.01 V and a scan rate of 50 mV·s-1. The solutions
were made in phosphate buffer 50 mM (Merck, Darmstadt, Germany) at pH=7.4 with
100 mM KCl as electrolyte and measured at room temperature without stirring.
Electrochemical Impedance Spectroscopy (EIS) was performed on an AUTOLAB
PGSTAT30 (Ecochemie, Netherlands), equipped with FRA module and with equivalent
set-up as in the voltammetric measurements.
7
3 Results and discussion
3.1 Chemical and morphological analysis of CDs
All the synthesized modified CDs where analysed by elementary analysis in order to
quantify the sulphur and nitrogen compositions (Table 1). The results confirm the
correct modification of the graphitic carbon dots, showing an appropriate doping grade,
around 10%; moreover the table demonstrates the previous absence of S and N in the
pristine material, as it corresponds.
By SEM analysis the morphology of the obtained N- and S-CDs (Figure 1) could be
visualized; the obtained images exhibit spherical shape and different degree of
aggregations for N-CDs (Figure 1B) and S-CDs (Figure 1C) compared with the raw CDs
(Figure 1A). This fact can be explained by the different surface charge of modified carbon
dots; S-CDs surface is populated by a higher concentration of sulphonic (R-SO3-) groups
and N-CDs by amide/amine (-CONH2/-CNH2) protonable groups.
The size histograms, made from different images and for approximately 150
measurements for each material, show mean diameters equal to 18, 16 and 4 nm for
CDs, N-CDs and S-CDs, respectively.
< Figure 1>
To support the stated doping, XPS analysis were performed with both N-CDs and S-CDs.
Figure 2 visualizes the different functional groups presented in both nano-dots. The
main contribution in both nanoparticles, is assigned to C 1s composition and attributed
8
to C-C / C-H at 284. 9 eV (~83 %), with small contributions related to C-O at 286 eV and
carbonyl and carboxylic groups at 287.3 eV. N-CDs and S-CDs showed the presence of
amine/amide (NH2/O=C-NH2) at 400.2 eV for N 1s contribution (Figure 2A) and
sulphate/sulphonic (SO42- (33%)/R-SO3- (67%)) functional groups at 168.2 and 169.8 eV
respectively, for S 2p contribution (Figure 2B). O 1s signals showed two main
contributions for C-O and O-C=O for N-CDs and an additional contribution for S=O in the
case of S-CDs.
< Figure 2>
Mass spectra of N- and S-CDs have confirmed the composition of both nanoparticles,
and they are presented in Figure 3. Detected signals, their position and their identity are
given in Table 2. Presented spectra are acquired by matrix-free approach, which was
possible due to optical and semi-conductive properties of carbon dots. Signals were
confirmed also by the spectra measured with organic matrices (data not shown), and
according to the mechanism of information, in the case of S-carbon dots, the addition
of protons and sodium/potassium was necessary for charge compensation of SO3- group
on the surface of CDs and subsequent ionization. Less additional positive ions were
required for the ionization of N-CDs. The number of carbon atoms was between 9 and
29, whereas in the spectra of S-CDs, a dimmer and a trimmer were detectable (at m/z
591.9 and at 887.9, respectively). They are most likely generated in the gas phase after
the laser induced desorption and ionization, as detected for other analytes [50].
< Figure 3>
9
3.2 Electrochemical characterization
Since the CDs are highly conductive, they were tested and characterized onto electrode
surfaces for they use as key transducer modifier material. The platform chosen for the
performance of electric measurement was the glassy carbon electrodes (GC) owing to
their well-known characteristics and responses. The placement of CDs on the surface of
these electrodes was carried out using polystyrene inks. The herein work has as
predecessor article with pristine carbon CDs developed in the same group [41]. This
study also used GC electrodes, which were modified via a drop casting method; in the
present work, due to the water solubility of N-CDs and S-CDs, it was mandatory the use
of an alternate method. For this reason, a conductive ink approach was attempted. The
employed ink had a basis of polystyrene as polymeric part, graphite as bulk conductive
material and mesitylene as solvent. To that suspension it was added a 10% w/w of
modifier, in this case the CDs.
When XPS was used to inspect, not the CDs themselves, but the ink surfaces, prepared
with the different doped and pristine CDs, no clear differences could be observed in the
spectra, suggesting the doping atoms are embedded in the graphitic core; in the same
sense, initial electrochemical evaluations were lacking clear signal and differentiation
between the different variants, suggesting us an activation of the deposited ink was
necessary in order to expose and make available the reactive spots to the
electrochemical processes.
On Figure 4A, it can be observed the effect of the activation process, where increasing
peaks appear and stabilize after ca. 10 cycles. In the same Figure, it also appears the
signal of a bare GC electrode which did not present any major change. Additionally, the
10
activation step was followed by electrochemical impedance spectroscopy (EIS); the
measurements were carried out in [Fe(CN)6]3-/4- between 0.5 MHz and 0.05 Hz with a
sinusoidal voltage perturbation of 10 mV amplitude. In Figures 4B and 4C, the EIS results
for non activated and activated electrodes, respectively, are depicted. The Nyquist plots
state equivalent results to those from cyclic voltammetry, since non-activated inks show
a larger charge transfer resistance than the activated electrodes. Moreover, a bare GC
electrode was measured in order to ensure the obtained results depend on the
activation of the ink and not on any alteration on the GC electrode.
< Figure 4>
Given the depositing process is manually conducted, it was necessary to study its
reproducibility and repeatability. For this, the active surface area of each electrode was
calculated with the Randles-Sevcik equation and with use of the [Fe(CN)6]3-/4-redox pair.
By applying 6 different voltammetry scan rates (10, 20, 50, 100, 200 and 300 mV·s-1) it
was obtained the following active areas (confidence interval calculated for 5 different
days and 5 different depositions): 42.18±0.06 mm2 for raw CDs ink, 29.19±0.07 mm2 for
S-CDs ink, 43.94±0.08 mm2 for N-CDs ink, 30.62±0.08 mm2 for graphite ink and
15.24±0.03 mm2 for the bare GC electrode (geometrical diameter, Ø=3 mm in all cases).
The results demonstrate a high reproducibility of deposition and activation, 0.26% in the
worst case, showing it is a robust modification technique. On the other hand, the
activation step demonstrates a noticeable increase of the active area (ca. a factor of 3)
in front of the bare GC electrode, leading this to higher current responses, markedly for
11
the N-doped CDs. In the individual use of these produced electrodes, their specific
reproducibility and repeatability values are 9.5% and 1%, respectively.
3.3 Electrochemical measurements
Once a satisfactory method for integrating and immobilizing the CD modifiers was
established, electrochemical measurements with different families of oxidizable organic
compounds were performed. Specifically, the chosen analytes were Ascorbic acid (AA),
Cysteine (CYS) and Salicylic Acid (SA), compounds that may be present in pharmaceutical
compositions and also be used as food additives [51-53]. The complexity to analyse them
via standard electrochemical methods, due to extreme oxidation potentials and low
defined peaks [54-56], makes of these three compounds interesting focus for study.
The results obtained for AA, SA and CYS, are summarized on Table 3. The values shown
in the table indicate a clear improvement of the sensitivity with use of the CDs; the
slopes of the modified electrodes vs. the bare GC electrode demonstrate a twice
increase, most of the times. This signal increase can be also observed on Figure 5, where
the effect of specially the N-doped CDs demonstrates, in the SA oxidation, the
improvements achieved. CYS is the compound interacting mostly with the S-CD modifier,
probably through its own thiol moiety, presenting an increase of more than 3.5 times
the current obtained for the bare glassy carbon electrode. This also demonstrates a good
synergy between the inks and synthesized carbon dots. The better linearities of the
measurements performed corresponded to AA and SA but not for CYS, as it is deduced
from the correlation coefficients. Also, Table 3 allows observing that there are no other
differences between the currents for the electrodes modified with the distinct inks,
12
which lead us to affirm that the improvements in current are, in first instance, a result
of the increase of the active surface.
< Figure 5>
But obviously, it is not only the increase of the current the benefit pursued with the
modification of the carbon electrode; concretely, the catalytic effects on the potential
are also as interesting as a further improvement in voltammetric measurements. And
herein we can observe two of these examples. First on Figure 5A, it is evident that the
different modifiers decrease the potential for the oxidation of SA vs. the GCE, not only
with reduction of the potential maximum, but also with a better definition of the peak.
As shown on Figure 5C, and more clearly depicted on Figure 6, a clear shift towards lower
potential is observed in the case of the CD modified electrodes for AA determination.
Particularly, the potentials are lowered ca. 200 mV, a relevant value. It is also important
to notice the special effect on the AA oxidation signal, where the peak is unfolded in two
overlapped peaks, which means the two different oxidations are better differentiated.
Finally, Figure 7 shows a complete calibration, in this case corresponding to SA with the
different electrodes. In this case, a higher signal with the pristine CDs is manifested,
while comparable responses for the doped and standard graphite inks seem confirmed;
consider here that the enhancement to remark with SA are the catalytic effects,
displayed in the reduction of the potential of the oxidation peaks, as commented above.
< Figure 6>
13
< Figure 7>
A final reflection to be done with the scrutiny of the signals obtained, is that, as it is
easily observed on the graphs, each modification produces different signal profiles, (this
is mostly evident with the SA and AA cases) making it an exceptional property for a
future application of carbon dots in studies of the electronic tongue type. Recalling the
electronic tongue definition, this is “an analytical instrument comprising an electrode
array of nonspecific, low-selective, stable and cross-sensitivity to different species” [57,
58]. The nature of the signals obtained lead us to regard the developed materials as
interesting transducer modifiers applied as sensors forming the array on electronic
tongues, as the starting point for its development is that the different sensors must
produce differentiated signals for the different compounds studied. This point is clearly
evident from the Figure 5, what proves the cross-sensitivity feature, and makes this a
starting point to consider further work on this direction.
4 Conclusions
N- and S-doped carbon dots were successfully synthesized and characterized via SEM
microscopy, MALDI TOF mass spectrometry, elementary analysis and XPS, in exploration
of new nanotechnology-type modifiers for voltammetry. Synthesized N- and S- CDs were
with regular shapes of nanoscale dimensions, and the modification of CDs was clearly
shown by elemental analysis, mass spectrometry and XPS. In this respect, we have
clearly demonstrated the potential of CDs and its derivates to catalyse the oxidation
potential and the ability to provide different voltammetric profiles for the same
14
compound, making them interesting and novel for their use in sensor arrays for
electronic tongues.
The data show a reproducible modification method with use of a conductive ink-like
preparation. The developed modification method is related with a remarkable active
surface area which produces a clear enhancement of the current signal. Overall, the
materials employed represent a remarkably green approach for obtaining and modifying
electrochemical sensors. Tested compounds with these CDs inks modified electrodes
were ascorbic acid, salicylic acid and cysteine, as generic oxidizable organic substances.
All the compounds examined showed signal improvement in their cyclic voltammetric
measurements. For example, signals of salicylic acid were shifted to lower potentials,
what is a proof of the increase of the heterogeneous electron transfer rate and
demonstrates an enhanced catalytic response. When comparing the different materials,
the effect of the doping heteroatom on the response is more noticeable for N-CDs than
for S-CDs, but both displayed better responses than the unmodified CDs, resulting on a
two-fold increase of the response of the GCE.
Acknowledgments
The authors thank to the project CTQ2016-80170-P (MINECO, Spain). ARDITI-Agência
Regional para o Desenvolvimento da Investigação Tecnologia e Inovação is also
acknowledged through the project M1420-01-0145-FEDER-000005-CQM+ (Madeira 14-
20 Program). MdV thanks the support from program ICREA Academia. MBSE thanks to
la Secretaria d’Universitats i Recerca del Departament d’Empreses i Coneixement de la
15
Generalitat de Catalunya and to European Social Fund, European Union for a FI
fellowship.
References
[1] X. Xu, R. Ray, Y.L. Gu, H.J. Ploehn, L. Gearheart, K. Raker, W. Scrivens, Electrophoretic
analysis and purification of fluorescent single-walled carbon nanotube fragments, J.
Am. Chem. Soc. 126 (2004) 12736-12737.
[2] J.C.G. Esteves da Silva, H.M.R. Goncalves, Analytical and bioanalytical applications of
carbon dots, TRAC-Trends Anal. Chem. 30 (2011) 1327-1336.
[3] H. Li, Z. Zang, Y. Liu, S.T. Lee, Carbon nanodots: synthesis, properties and
applications, J. Mater. Chem. 22 (2012) 24230-24253.
[4] X. Sun, Y. Lei, Fluorescent carbon dots and their sensing applications, TrAC Trends in
Anal. Chem. 89 (2017) 163-180.
[5] S. Zhu, Q. Meng, L. Wang, Z. Zhang, S. Song, J. Jin, Z. Zhang, S. Sun, W. Wang, Y. Yang,
Highly photoluminescent carbon dots for multicolor patterning, sensors, and
bioimaging, Angew. Chem. Int. Ed. 125 (2013) 4045-4049.
[6] A. Sachdev, P. Gopinath, Green synthesis of multifunctional carbon dots from
coriander leaves and their potential application as antioxidants, sensors and
bioimaging agents, Analyst 140 (2015) 4260-4269.
[7] W. Lu, X. Qin, S. Liu, G. Chang, Y. Zhang, Y. Luo, A.M. Asiri, A.O. Al-Youbi, X. Sun,
Economical, green synthesis of fluorescent carbon nanoparticles and their use as
probes for sensitive and selective detection of mercury(II) ions, Anal. Chem. 84
(2012) 5351-5357.
[8] M. Algarra, M. Pérez-Martín, M. Cifuentes-Rueda, J. Jiménez-Jiménez, J.E. da Silva, T.
Bandosz, E. Rodriguez-Castellón, J.T. López Navarrete, J. Casado, Carbon dots
obtained using hydrothermal treatment of formaldehyde. Cell imaging in vitro,
Nanoscale 6 (2014) 9071-9077.
[9] H. Shi, J. Wei, L. Qiang, X. Chen, X. Meng, Fluorescent carbon dots for bioimaging and
biosensing applications, J. Biomed. Nanotechnol. 10 (2014) 2677–2679.
16
[10] A.H. Loo, Z. Sofer, D. Bouša, P. Ulbrich, A. Bonanni, M. Pumera, Carboxylic Carbon
Quantum Dots as a Fluorescent Sensing Platform for DNA Detection, ACS Appl.
Mater. Interfaces. 8 (2016) 1951–1957.
[12] A. Rahy, C. Zhou, J. Zheng, S. Park, M.J. Ki, I. Jang, S.J. Cho, D.J. Yang,
Photoluminiscent carbon nanoparticles produced by confined combustion of
aromatic compounds, Carbon 50 (2012) 1298-1302.
[13] Q.L. Zhao, Z.L. Zhang, B.H. Huang, J. Peng, M. Zhang, D.W. Pang, Facile preparation of
low cytotoxicity fluorescent carbon nanocrystals by electrooxidation of graphite,
Chem. Comm. 41 (2008) 5116-5118.
[14] S.C. Ray, A. Saha, N.R. Jana, R. Sarkar, Fluorescent carbon nanoparticles: Synthesis,
characterization and bioimaging application, J. Phys. Chem. C 113 (2009) 18546-
18551.
[15] S. Li, Z. Guo, Y. Zhang, W. Xue, Z. Liu, Blood Compatibility Evaluations of Fluorescent
Carbon Dots, ACS Appl. Mater. Interfaces 7 (2015) 19153–19162.
[16] Y. Hou, Q. Lu, J. Deng, H. Li, Y. Zhang, One-pot electrochemical synthesis of
functionalized fluorescent carbon dots and their selective sensing for mercury ion,
Anal. Chim. Acta 866 (2015) 69-74.
[17] T.S. Atabaev. Doped carbon dots for sensing and bioimaging applications: A
minireview, Nanomaterials 8 (2018) 342.
[18] M. Zhou, Z. Zhou, A. Gong, Y. Zhang, Q. Li. Synthesis of highly photoluminiscent
carbon dots via citric acid and Tris for iron(III) ion sensors and bioimaging, Talanta
143 (2015) 107-113.
[19] Y. Du, S. Guo, Chemically doped fluorescent carbon and graphene quantum dots for
bioimaging, sensor, catalytic and photoelectronic applications, Nanoscale 8 (2016)
2532-2543.
[20] S. Chaudhary, A. Umar, K. K. Bhasin, S. Singh, Applications of carbon dots in
nanomedicine, J. Biomed. Nanotech. 13 (2017) 591-637.
[21] X. Lin, G. Gao, L. Zheng, Y. Chi, G. Chen. Encapsulation of strongly fluorescent carbon
quantum dots in metal-organic frameworks for enhancing chemical sensing, Anal.
Chem., 86 (2014), pp. 1223-1228.
17
[22] N. A. Travlou, D. A. Giannakoudakis, M. Algarra, A.M. Labella, E. Rodríguez-Castellón,
T. J. Bandosz, S- and N-doped carbon quantum dots: Surface chemistry dependent
antibacterial activity, Carbon 135 (2018) 104-111.
[23] J. Praneerad, N. Thongsai, P. Supchocksoonthorn, S. Kladsomboon, P. Paoprasert,
Multipurpose sensing applications of biocompatible radish-derived carbon dots as
Cu2+ and acetic acid vapor sensors, Spectrochim. Acta Part A Mol. Biomol. Spectrosc.
211 (2019) 59–70.
[24] Y. Chen, D. Gao, H. Bai, H. Liu, S. Lin, Y. Jiang, Carbon Dots and 9AA as a Binary Matrix
for the Detection of Small Molecules by Matrix-Assisted Laser Desorption/Ionization
Mass Spectrometry, J. Am. Soc. Mass Spectr. 27 (2016) 1227–1235.
[25] M. Shahnawaz Khan, M.L. Bhaisare, S. Pandey, A. Talib, S.-M. Wu, S.K. Kailasa, H.-F.
Wu, Exploring the ability of water soluble carbon dots as matrix for detecting
neurological disorders using MALDI-TOF MS, Int. J. Mass Spectrom. 393 (2015) 25–33.
[26] J. Zhou, X. Shan, J. Ma, Y. Gu, Z. Qian, J. Chen, H. Feng. Facile synthesis of P-doped
carbon quantum dots with highly efficient photoluminescence, RSC Advances 4
(2014) 5465-5468.
[27] Y. Dong, H. Pang, H.B. Yang, C. Guo, J. Shao, Y. Chi, C.M. Li, T. Yu, Carbon-based dots
co-doped with nitrogen and sulfur for high quantum yield and excitation-
independent emission, Angew. Chem. Int. Ed. 52 (2013) 7800-7804.
[28] Q. Xu, P. Pu, J. Zhao, C. Dong, C. Gao, Y. Chen, J. Chen, Y. Liu, H. Zhou. J. Mater. Chem.
A, Preparation of highly photoluminiscent sulfur-doped carbondots for Fe(III)
detection, J. Mater. Chem. A 3 (2015) 542-546.
[29] Z. Lin, J. Wu, Y. Dong, P. Xie, Y. Zhang, Z. Cai, Nitrogen and Sulfur Co-doped Carbon-
Dot-Assisted Laser Desorption/Ionization Time-of-Flight Mass Spectrometry Imaging
for Profiling Bisphenol S Distribution in Mouse Tissues, Anal. Chem. 90 (2018) 10872–
10880.
[30] Y. Jiang, B. Wang, F. Meng, Y. Cheng, C. Zhu, Microwave-assisted preparation of N-
doped carbon dots as a biosensor for electrochemical dopamine detection, J. Colloid.
Interface Sci. 452 (2015) 199-202.
18
[31] X. Teng, C. Ma, C. Ge, M. Yan, J. Yang, Y. Zhang, P. C. Morais, H. Bi. Green
synthesis of nitrogen-doped carbon dots from konjac flour with “off-on” fluorescence
by Fe3+ and L-lysine for bioimaging, J. Mater. Chem. B 2 (2014) 4631-4639.
[32] B.B. Campos, C. Abellán, M. Zougagh, J. Jiménez-Jiménez, E. Rodríguez-Castellón,
J.C.G. Esteves da Silva, A. Ríos, M. Algarra, Fluorescent chemosensor for pyridine
based on N-doped carbon dots, Colloid. Interface Sci. 458 (2015) 209-216.
[33] H. Liu, Y. Zhang, J. H. Liu, C.Z. Huang, P. Hou, J. Zhou, C.Z. Huang, Preparation of
nitrogen-doped carbon dots with high quantum yield from Bombyx mori silk for
Fe(III) ion detection, RSC Advances 7 (2017) 50584-50590.
[34] M.Y. Emran, M.A. Shenashena A.A. Abdelwahab, M. Abdelmottaleb, S.A. El-Safty,
Facile synthesis of microporous sulfur-doped carbon spheres as electrodes for
ultrasensitive detection of ascorbic acid in food and pharmaceutical products, New J.
Chem. 42 (2018) 5037-5044.
[35] H. Liu, Y. Zhang, C. Huang, Development of nitrogen and sulfur-doped carbon dots for
cellular imaging, J. Pharm. Anal. 9 (2019) 127-132.
[36] Y. Wang, L. Yan, G. Ji, C. Wang, H. Gu, Q. Luo, Q. Chen , L. Chen Y. Yang, C. Q. Ma, X.
Liu, Synthesis of N,S-doped carbon quantum dots for use in organic solar cells as the
ZnO modifier to eliminate the light-soaking effect, ACS Appl. Mater. Interfaces 11
(2019) 2243-2253.
[37] C. Hu, M. Li, J. Qiu, Y.P. Sun, Design and fabrication of carbon dots for energy
conversion and storage, Chem. Soc. Rev. 48 (2019) 2315-2337.
[38] S. Campuzano, P. Yáñez-Sedeño, J.M. Pingarrón, Carbon dots and graphene quantum
dots in electrochemical biosensing , Nanomaterials, 9 (2019) 634.
[39] M. Li, T. Chen, J.J. Gooding, J. Liu, A review of carbon and graphene quantum dots for
sensing, ACS Sensors 4 (2019) 1732-1748.
[40] H.V. Nguyen, L. Richtera, A. Moulick, K. Xhaxhiu, J. Kudr, N. Cernei, H. Polanska, Z.
Heger, M. Masarik, P. Kopel, M. Stiborova, T. Eckschlager, V. Adam, R. Kizek,
Electrochemical sensing of etoposide using carbon quantum dot modified glassy
carbon electrode, Analyst 141 (2016) 2665-2675.
[41] M. Algarra, A. González-Calabuig, K. Radotić, D. Mutavdzic, C.O. Ania, J.M. Lázaro-
Martínez, J. Jiménez-Jiménez, E. Rodríguez-Castellón, M. del Valle, Enhanced
19
electrochemical response of carbon quantum dot modified electrodes, Talanta, 178
(2018) 679–685.
[42] J.Tang, X. Ma, J. Liu, S. Zheng, J. Wang, Simultaneous determination of hydroquinone
and catechol using carbon glass electrode modified with graphene quantum dots, Int.
J. Electrochem. Sci. 13 (2018) 11250-11262.
[43] S.S. Shankar, R.M. Shereema, V. Ramachandran, T.V. Sruthi, V.B.S. Kumar, R.B. Rakhi,
Carbon Quantum Dot-Modified Carbon Paste Electrode-Based Sensor for Selective
and Sensitive Determination of Adrenaline, ACS Omega, 4 (2019) 7903-7910.
[44] S. Baluta, A. Lesiak, J. Cabaj, Graphene Quantum Dots-based electrochemical
biosensor for catecholamine neurotransmitters detection, Electroanalysis 30 (2018)
1773-1782.
[45] W. Chen, W. Weng, X. Niu, X. Li, Y. Men, W. Sun, G. Li, L. Dong, Boron-doped
Graphene quantum dots modified electrode for electrochemistry and electrocatalysis
of haemoglobin, J. Electroanal. Chem. 823 (2018) 137–145.
[46] G. Muthusankar, R. Sasikumar, S.-M. Chen, G. Gopu, N. Sengottuvelan, S.-P. Rwei,
Electrochemical synthesis of nitrogen-doped carbon quantum dots decorated copper
oxide for the sensitive and selective detection of non-steroidal anti-inflammatory
drug in berries, J. Colloid Interface Sci. 523 (2018) 191–200.
[47] L. Fu, A. Wang, G. Lai, C.-T. Lin, J. Yu, A. Yu, Z. Liu, K. Xie, W. Su, A glassy carbon
electrode modified with N-doped carbon dots for improved detection of hydrogen
peroxide and paracetamol, Microchimica Acta 185 (2018) 87.
[48] M.I. González-Sánchez, B. Gómez-Monedero, J. Agrisuelas, J. Iniesta, E. Valero, Highly
activated screen-printed carbon electrodes by electrochemical treatment with
hydrogen peroxide, Electrochem. Commun. 91 (2018) 36–40.
[49] J. Wang, M. Pedrero, H. Sakslund, O. Hammerich, J. Pingarron, Electrochemical
activation of screen-printed carbon strips, Analyst 121 (1996) 345-350.
[50] M. Petković, J. Schiller, M. Müller, R. Süß, K. Arnold, J. Arnhold, Detection of Adducts
with Matrix Clusters in the Positive and Negative Ion Mode MALDI-TOF Mass Spectra
of Phospholipids, Z. Naturforsch. B. 64 (2009) 331–334.
[51] J.F. Gregory, III, Ascorbic Acid Bioavailability in Foods and Supplements, Nutr. Rev. 51
(2009) 301–303.
20
[52] S. Schwimmer, D.G. Guadagni, Cysteine induced odor intensification in onions and
other foods, J. Food Sci. 32 (1967) 405–408.
[53] J. Radtke, J. Linseisen, G. Wolfram, Phenolsäurezufuhr Erwachsener in einem
bayerischen Teilkollektiv der Nationalen Verzehrsstudie, Eur. J. Nutr. 37 (1998) 190–
197.
[54] F. Cao, Y. Huang, F. Wang, D. Kwak, Q. Dong, D. Song, J. Zeng, Y. Lei, A high-
performance electrochemical sensor for biologically meaningful l-cytokine based on a
new nanostructured l-cytokine electrocatalyst, Anal. Chim. Acta 1019 (2018) 103–
110.
[55] M. Bonet-San-Emeterio, A. Gonzalez, M. del Valle, Artificial neural networks for the
resolution of dopamine and serotonin complex mixtures using a graphene-modified
carbon electrode, Electroanalysis 31 (2019) 390-397.
[56] D. Evans, J.P. Hart, G. Rees, Voltammetric behaviour of salicylic acid at a glassy
carbon electrode and its determination in serum using liquid chromatography with
amperometric detection, Analyst 116 (1991) 803-806.
[57] Y. Vlasov, A. Legin, A. Rudnitskaya, C. Di Natale, A. D’Amico, Nonspecific sensor arrays
(electronic tongue) for chemical analysis of liquids (IUPAC Technical Report), Pure
Appl. Chem. 77 (2005) 1965–1983.
[58] M. del Valle, Bioinspired sensor systems, Sensors 11 (2011) 10180-10186.
21
Table 1. Elementary analysis (weight %) of the nanoparticles prepared in this study a.
%C %H %N %S
S-CDs 34.72 ± 0.34 3.88 ± 0.23 - 13.22 ± 0.22
N-CDs 54.98 ± 0.05 7.50 ± 0.51 10.72 ± 0.02 -
a
Average value ± SD (n=3).
22
Table 2. The identity and position (m/z) of signals detected in the LDI TOS mass spectra
of N-,S-carbon dots. All detected signals correspond to the singly-positively charged
ions.
Position, m/z Identity
191.9 C9SO3+4H
227.9 C12SO3+4H
261.9 C13SO3+3H+Na
281.1 C16(CONH2)2+H
283.9 C13SO3+2H+2Na
292.1 C17(CONH2)2+H
303.1 C16(CONH2)2+Na
305.9 C13SO3++H+3Na
327.9 C13SO3+4Na
331.9 C17(CONH2)2+K
339.9 C14SO3+4Na
389.9 C13(SO3)2+5H+5Na
443.9 C13(SO3)2+4H+2Na+2K
558.1 C28(CONH2)4-2H+2Na
591.9 (C13SO3+2H+2Na)2
635.1 C29(CONH2)6-+Na
887.9 (C13SO3+2H+2Na)3
23
Table 3. Summary table of calibration experiments for AA, SA and Cys for bare GC,
graphite ink, C-CDs, N-CDs and S-CDs electrodes. All measurement carried out at same
conditions and same range of concentrations.
GC Electrode Graphite C-CDs N-CDs S-CDs
Ink
Ascorbic Acid
Peak potential 0.32 0.32 0.32 0.42 0.32
(V)
Calibration line y=1.82·10-2 y=3.24·10-2 y=2.62·10-2 y=2.39·10-2 y=2.07·10-2
x+4.26·10-7 x+2.79·10-6 x+1.24·10-6 x+2.61·10-6 x+2.13·10-6
R2 0.9995 0.9996 0.9999 0.9993 0.9997
LOD (µM) 101 88.7 20.7 121 78.0
Salicylic Acid
Peak potential 0.91 0.84 0.84 0.87 0.85
(V)
Calibration line y=2.15·10-2 y=3.55·10-2 y=4.64·10-2 y=3.85·10-2 y=3.60·10-2
x+1.12·10-6 x+2.72·10-6 x+4.94·10-6 x+3.34·10-6 x+3.04·10-6
R2 0.9995 0.998 0.998 0.998 0.998
LOD (µM) 51.1 113 101 91.2 94.8
Cysteine
Peak potential 0.64 0.64 0.63 0.64 0.63
(V)
Calibration line y=5.60·10-3 y=2.01·10-2 y=1.73·10-2 y=1.43·10-2 y=1.70·10-2
x+9.03·10-7 x+9.33·10-6 x+7.31·10-6 x+7.29·10-6 x+8.27·10-6
R2 0.987 0.970 0.970 0.950 0.972
LOD (µM) 287 420 427 548 407
24
CAPTIONS FOR FIGURES
Figure 1. SEM images and their size histograms of (A) CDs and its doped derivates: (B)
N-CDs and (C) S-CDs.
Figure 2. Percentage of different bond types observed in the XPS analysis of the
functional groups contributions for the N-CDs, (A), and the S-CDs, (B).
Figure 3. Positive ion laser induced desorption and ionization time-of-flight mass
spectrometry of N- (a) and S-(b) doped CDs. The spectra are acquired in reflector mode
under delayed extraction conditions. Each spectrum represents the average of 2000
individual laser shots at the laser frequency of 200 Hz.
Figure 4. Electrochemical characterization of modified GC electrodes, (A) reduction in
H2O2 solution of the 5 studied electrodes, B and C depicts the Nyquist plots measured
before the activation (B) and after the activation (C), both of them in [Fe(CN)6]3-/4-
solution at pH=7.0.
Figure 5. Comparison of the cyclic voltammograms, at the same concentration of
analyte, obtained from the 5 considered electrodes for (A) SA, (B) AA and (C) Cys.
Figure 6. Cyclic voltammograms obtained with AA when a N-CDs modified electrode is
used for the measurement of different concentrations (0.002 to 3 mM) of the analyte.
Figure 7. Calibration curves for SA measured with GC electrode and Graphite ink, C-CDs,
N-CDs and S-CDs GC modified electrodes. Error bars corresponding to three calibrations,
made in different days.