Calibration of low-temperature ac susceptometers with a copper cylinder standard
Chen, Du-Xing (Universitat Autònoma de Barcelona. Departament de Física)
Skumryev, Vassil 
(Universitat Autònoma de Barcelona. Departament de Física)
| Fecha: |
2010 |
| Resumen: |
A high-quality low-temperature ac susceptometer is calibrated by comparing the measured ac susceptibility of a copper cylinder with its eddy-current ac susceptibility accurately calculated. Different from conventional calibration techniques that compare the measured results with the known property of a standard sample at certain fixed temperature T, field amplitude Hm, and frequency f, to get a magnitude correction factor, here, the electromagnetic properties of the copper cylinder are unknown and are determined during the calibration of the ac susceptometer in the entire T, Hm, and f range. It is shown that the maximum magnitude error and the maximum phase error of the susceptometer are less than 0. 7% and 0. 3°, respectively, in the region T = 5-300 K and f = 111-1111 Hz at Hm = 800 A/m, after a magnitude correction by a constant factor as done in a conventional calibration. However, the magnitude and phase errors can reach 2% and 4. 3° at 10 000 and 11 Hz, respectively. Since the errors are reproducible, a large portion of them may be further corrected after a calibration, the procedure for which is given. Conceptual discussions concerning the error sources, comparison with other calibration methods, and applications of ac susceptibility techniques are presented. |
| Nota: |
This article may be downloaded for personal use only. Any other use requires prior permission of the author and the American Institute of Physics. |
| Derechos: |
Aquest material està protegit per drets d'autor i/o drets afins. Podeu utilitzar aquest material en funció del que permet la legislació de drets d'autor i drets afins d'aplicació al vostre cas. Per a d'altres usos heu d'obtenir permís del(s) titular(s) de drets.  |
| Lengua: |
Anglès |
| Documento: |
Article ; recerca ; Versió publicada |
| Materia: |
Calibratge ;
Coure ;
Materials ;
Propietats magnètiques |
| Publicado en: |
Review of scientific instruments, Vol. 81, Núm. 2 (Feb. 2010) , p. 025104/1-025104/10, ISSN 0034-6748 |
DOI: 10.1063/1.3309779
El registro aparece en las colecciones:
Documentos de investigación >
Documentos de los grupos de investigación de la UAB >
Centros y grupos de investigación (producción científica) >
Ciencias >
Grupo de nanoingeniería de materiales, nanomagnetismo y nanomecánica (Gnm3)Artículos >
Artículos de investigaciónArtículos >
Artículos publicados
Registro creado el 2013-06-25, última modificación el 2024-11-30