Web of Science: 2 citations, Scopus: 4 citations, Google Scholar: citations
Optimal clinical time for reliable measurement of transcutaneous CO2 with ear probes : counterbalancing overshoot and the vasodilatation effect
Domingo, Christian (Universitat Autònoma de Barcelona. Departament de Medicina)
Canturri, Elisa (Universitat Autònoma de Barcelona. Departament de Medicina)
Moreno Gomez de Bonilla, Amalia (Universitat Autònoma de Barcelona. Departament de Medicina)
Espuelas, Humildad (Universitat Autònoma de Barcelona. Departament de Medicina)
Vigil Giménez, Laura (Universitat Autònoma de Barcelona. Departament de Medicina)
Luján, Manel (Universitat Autònoma de Barcelona. Departament de Medicina)

Date: 2010
Abstract: Objectives: To determine the optimal clinical reading time for the transcutaneous measurement of oxygen saturation (SpO2) and transcutaneous CO2 (TcPCO2) in awake spontaneously breathing individuals, considering the overshoot phenomenon (transient overestimation of arterial PaCO2). Experimental section: Observational study of 91 (75 men) individuals undergoing forced spirometry, measurement of SpO2 and TcPCO2 with the SenTec monitor every two minutes until minute 20 and arterial blood gas (ABG) analysis. Overshoot severity: (a) mild (0. 1-1. 9 mm Hg); (b) moderate (2-4. 9 mm Hg); (c) severe: (>5 mm Hg). The mean difference was calculated for SpO2 and TcPCO2 and arterial values of PaCO2 and SpO2. The intraclass correlation coefficient (ICC) between monitor readings and blood values was calculated as a measure of agreement. RESULTS: The mean age was 63. 1 ± 11. 8 years. Spirometric values: FVC: 75. 4 ± 6. 2%; FEV1: 72. 9 ± 23. 9%; FEV1/FVC: 70 ± 15. 5%. ABG: PaO2: 82. 6 ± 13. 2; PaCO2: 39. 9. 1 ± 4. 8 mmHg; SaO2: 95. 3 ± 4. 4%. Overshoot analysis: overshoot was mild in 33 (36. 3%) patients, moderate in 20 (22%) and severe in nine (10%); no overshoot was observed in 29 (31%) patients. The lowest mean differences between arterial blood gas and TcPCO2 was -0. 57 mmHg at minute 10, although the highest ICC was obtained at minutes 12 and 14 (>0. 8). The overshoot lost its influence after minute 12. For SpO2, measurements were were reliable at minute 2. CONCLUSIONS: The optimal clinical reading measurement recommended for the ear lobe TcPCO2 measurement ranges between minute 12 and 14. The SpO2 measurement can be performed at minute 2.
Rights: Aquest document està subjecte a una llicència d'ús Creative Commons. Es permet la reproducció total o parcial, la distribució, la comunicació pública de l'obra i la creació d'obres derivades, fins i tot amb finalitats comercials, sempre i quan es reconegui l'autoria de l'obra original. Creative Commons
Language: Anglès
Document: Article ; recerca ; Versió publicada
Subject: Transcutaneous CO2 ; Optimal reading time ; Overshoot phenomenon
Published in: Sensors (Basel, Switzerland), Vol. 10 (2010) , p. 491-500, ISSN 1424-8220

DOI: 10.3390/s100100491
PMID: 22315552


10 p, 63.1 KB

The record appears in these collections:
Articles > Research articles
Articles > Published articles

 Record created 2016-03-04, last modified 2024-04-29