Web of Science: 35 citas, Scopus: 37 citas, Google Scholar: citas,
ADAPT technique with ACE68 and ACE64 reperfusion catheters in ischemic stroke treatment : Results from the PROMISE study
Schramm, P. (Department of Neuroradiology. University Medical Center Schleswig-Holstein)
Navia, P. (Hospital Universitario de Donostia (Sant Sebastià, País Basc))
Papa, R. (Radiology. Universita Degli Studi di Messina)
Zamarro, J. (Hospital Universitario Virgen de la Arrixaca (Múrcia))
Tomasello, A. (Hospital Universitari Vall d'Hebron. Institut de Recerca)
Weber, W. (Radiology and Neuroradiology. Ruhr-University Bochum. University Medical Center Langendreer. Bochum. Germany)
Fiehler, J. (Department of Neuroradiology. University Medical Center Hamburg Eppendorf. Hamburg. Germany)
Michel, P. (University of Lausanne)
Pereira, Vitor Mendes (Medical Imaging. Toronto Western Hospital)
Krings, T. (Division of Neurosurgery. Department of Surgery Toronto Western Hospital. University Health Network)
Gralla, J. (Department for Diagnostic and Interventional Neuroradiology. Inselspital University of Bern)
Santalucia, P. (IRCCS Centro Neurolesi Bonino Pulejo. Piemonte Hospital)
Pierot, L. (Radiology. University Hospital Reims)
Lo, T. H. (Department of Interventional Radiology and Interventional Neuroradiology. Universitair Medisch Centrum Utrecht)
Universitat Autònoma de Barcelona

Fecha: 2019
Resumen: Background and purpose The recent randomized trials demonstrated the benefit of mechanical thrombectomy in stroke therapy. However, treatment using different strategies is an ongoing area of investigation. The PROMISE study analyzed the safety and effectiveness of the Penumbra System with the ACE68 and ACE64 reperfusion catheters in aspiration thrombectomy of stroke, using A Direct Aspiration First Pass Technique (ADAPT). Methods PROMISE was a prospective study which enrolled 204 patients with intracranial anterior circulation large vessel occlusion (LVO) ischemic stroke in 20 centers from February 2016 to May 2017. Initial treatment was with the ACE68/ACE64 catheters within 6 hours of symptom onset. Imaging and safety review was performed by an independent Core Laboratory and a Clinical Events Committee. The primary angiographic outcome was revascularization to mTICI 2b-3 at immediate post-procedure and the primary clinical outcome was 90-day modified Rankin Scale (mRS) score ≤2. Safety assessment included device-and procedure-related serious adverse events (SAEs), symptomatic intracranial hemorrhage (sICH), mortality, and embolization of new territory (ENT). Results Enrolled patients had a median age of 74 (IQR 65-80) years and a median admission NIHSS of 16 (IQR 11-20). The post-procedure mTICI 2b-3 revascularization rate was 93. 1% and the 90-day mRS 0-2 rate was 61%. Device-and procedure-related SAEs at 24 hours occurred in 1. 5% and 3. 4%, respectively, 90-day mortality was 7. 5%, sICH occurred in 2. 9% while ENT occurred in 1. 5%. Conclusions For frontline therapy of LVO stroke, the ACE68/ACE64 catheters for aspiration thrombectomy were found to be safe and showed similar efficacy to randomized trials using other revascularization techniques. Clinical Trial Registration NCT02678169; Pre-results.
Nota: This study was funded by Penumbra, inc (9508).
Nota: Competing interests The following investigators report financial conflicts with this study: Ps reports being a consultant for Penumbra and stryker, and receiving research support from Penumbra, Philips, and siemens. rP reports proctoring and consulting for Medtronic, Penumbra, and implemed. WW reports a proctor agreement with Microvention/Terumo, Medtronic, and Phenox, and being a consultant with Penumbra (speaker and chairman fees). JF reports being a consultant for Penumbra, acandis, Boehringer ingelheim, cerenovus, covidien, Medtronic, Microvention, route92, and stryker, and receiving research support from the german Ministry of science and education (BMBF), german Ministry of economy and innovation (BMWi), german research Foundation (DFg), european Union (eU), hamburgische investitions-und Förderbank (iFB), Medtronic, Microvention, Philips, and stryker. PM reports speaker and consulting fees from Medtronic, used for stroke education. TK and VMP received funding from the sponsor for imaging core lab oratory activities. VMP reports being a consultant for Penumbra (PrOMise study), Medtronic, stryker, and neurovasc. lP received funding from Penumbra for cec activities and is a consultant for Balt, cerenovus/neuravi, and Microvention.
Derechos: Aquest document està subjecte a una llicència d'ús Creative Commons. Es permet la reproducció total o parcial, la distribució, la comunicació pública de l'obra i la creació d'obres derivades, sempre que no sigui amb finalitats comercials, i sempre que es reconegui l'autoria de l'obra original. Creative Commons
Lengua: Anglès
Documento: Article ; recerca ; Versió publicada
Publicado en: Journal of Neurointerventional Surgery, Vol. 11 Núm. 3 (march 2019) , p. 226-231, ISSN 1759-8486

DOI: 10.1136/neurintsurg-2018-014122
PMID: 30061367


6 p, 372.7 KB

El registro aparece en las colecciones:
Artículos > Artículos de investigación
Artículos > Artículos publicados

 Registro creado el 2020-06-03, última modificación el 2022-07-30



   Favorit i Compartir