Web of Science: 1 citas, Scopus: 1 citas, Google Scholar: citas
Impact of the dispersive patch placement on dissipated power in radiofrequency ablation for pulmonary vein isolation via a virtual patient study
Anees, M. (Johann Radon Institute for Computational and Applied Mathematics (RICAM). Austrian Academy of Sciences)
Moreno Weidmann, Zoraida (Institut de Recerca Sant Pau)
Viladés Medel, David (Institut de Recerca Sant Pau)
Guerra Ramos, José María (Institut de Recerca Sant Pau)
Gerardo-Giorda, L. (Johannes Kepler University)
Petras, A. (Austrian Academy of Sciences)
Universitat Autònoma de Barcelona

Fecha: 2025
Resumen: Radiofrequency ablation (RFA) is a minimally invasive technique for treating arrhythmias by interrupting abnormal electrical signals in the heart. Through a catheter tip, it delivers an alternating current that flows through the heart muscle tissue and the blood to a dispersive patch on the patient's skin. This study aims to test the hypothesis that the placement of the dispersive patch affects the efficacy and safety of RFA. By optimizing the patch position, the procedure could be made more effective and less risky for patients. A 3D in-silico model, based on patient imaging data, was developed to examine the effects of dispersive patch (DP) positioning on electric field distribution within cardiac tissue and the torso during RFA. We conducted 80 computer simulations using a CT-segmented torso model, exploring various DP and electrode configurations while applying standard (25 W) and high (90 W) power settings. For each configuration, we assessed the effectiveness of the DP in delivering power to cardiac tissue near the electrode. The main finding indicates that DP efficacy is significantly influenced by the current delivered to cardiac tissue. Notably, using an anterior patch during ablation proved more effective for the posterior left atrium compared to a posterior patch.
Derechos: Aquest document està subjecte a una llicència d'ús Creative Commons. Es permet la reproducció total o parcial, la distribució, la comunicació pública de l'obra i la creació d'obres derivades, fins i tot amb finalitats comercials, sempre i quan es reconegui l'autoria de l'obra original. Creative Commons
Lengua: Anglès
Documento: Article ; recerca ; Versió publicada
Materia: Atrial Fibrillation ; Catheter Ablation ; Computer Simulation ; Electrodes ; Humans ; Pulmonary Veins ; Radiofrequency Ablation ; Tomography, X-Ray Computed
Publicado en: Scientific reports, Vol. 15 Núm. 1 (december 2025) , p. 6986, ISSN 2045-2322

DOI: 10.1038/s41598-025-90158-1
PMID: 40011542


11 p, 2.5 MB

El registro aparece en las colecciones:
Documentos de investigación > Documentos de los grupos de investigación de la UAB > Centros y grupos de investigación (producción científica) > Ciencias de la salud y biociencias > Institut de Recerca Sant Pau
Artículos > Artículos de investigación
Artículos > Artículos publicados

 Registro creado el 2025-10-22, última modificación el 2026-01-01



   Favorit i Compartir