Pàgina inicial > Articles > Articles publicats > Assessment of a Novel Automatic Real-Time PCR Assay on the Cobas 4800 Analyzer as a Screening Platform for Hepatitis C Virus Genotyping in Clinical Practice : |
Data: | 2017 |
Resum: | The unequivocal identification of hepatitis C virus (HCV) subtypes 1a/1b and genotypes 2 to 6 is required for optimizing the effectiveness of interferon-free, direct-acting antiviral therapies. We compared the performance of a new real-time HCV genotyping assay used on the Cobas 4800 system (C4800) with that of high-resolution HCV subtyping (HRCS). In total, 502 samples were used, including 184 samples from chronic HCV patients (from routine laboratory activity during April 2016), 5 stored samples with double HCV genotype infections for testing the limitations of the method, and 313 samples from a screening protocol implemented in our hospital (from May to August 2016) based on the new method to further determine its genotyping accuracy. A total of 282 samples, including 171 from April 2016 (the 13 remaining had too low of a viral load for HRCS), 5 selected with double infections, and 106 from screening, were analyzed by both methods, and 220 were analyzed only by the C4800. The C4800 correctly subtyped 125 of 126 1a/1b samples, and the 1 remaining sample was reported as genotype 1. The C4800 correctly genotyped 38 of 45 non-1a/1b samples (classified by HRCS), and it reported the remaining 7 samples as indeterminate. One hundred two of 106 non-1a/1b genotype samples that were identified using the C4800 for screening were confirmed by HRCS. In the 4 remaining samples, 3 were correctly reported as genotype 1 (without defining the subtype) and 1 was reported as indeterminate. None of the samples were misgenotyped. Four of 7 samples with double HCV infections were correctly genotyped by the C4800. Excluding the 5 selected double-infected samples, the C4800 showed 95. 7% concordant results for genotyping HCVs 2 to 6 and 1a/1b subtyping, and 99. 2% concordance for subtyping 1a/1b single infections in clinical samples. To improve laboratory workflow, we propose using the C4800 as a first-line test for HCV genotyping and 1a/1b classification, followed by transferring non-1a/1b samples to a center where HRCS is available, if further characterization is needed. |
Ajuts: | Instituto de Salud Carlos III PI13-00456 Instituto de Salud Carlos III PI15-00829 Instituto de Salud Carlos III PI15-00856 Instituto de Salud Carlos III PI16-00337 Instituto de Salud Carlos III GLD14-00296 Ministerio de Economía y Competitividad IDI-20151125 |
Drets: | Aquest document està subjecte a una llicència d'ús Creative Commons. Es permet la reproducció total o parcial, la distribució, la comunicació pública de l'obra i la creació d'obres derivades, fins i tot amb finalitats comercials, sempre i quan es reconegui l'autoria de l'obra original. |
Llengua: | Anglès |
Document: | Article ; recerca ; Versió publicada |
Matèria: | Diagnostics ; Genotype identification ; Hepatitis C virus ; Molecular subtyping |
Publicat a: | Journal of Clinical Microbiology, Vol. 55 (january 2017) , p. 504-509, ISSN 1098-660X |